FORGOT YOUR DETAILS?

C-Reactive Protein (CRP) Test Cassette

Diagnostic Kit for the Quantitative Determination of C-reaction protein (Fluorescence Lateral Flow Immunoassay)

EU CE Mark. Not for home use.

SKU: INFL-003 Category:

Description

This kit is used for the in vitro quantitative detection of C-reactive protein (CRP) concentration in human serum/plasma and whole blood samples.
CRP is a typical acute phase protein. It is synthesized by the liver and the epithelial cell in response to infections or tissue damages, and the synthesis is triggered by interleukin-6 (IL-6) and other cytokines produced by macrophages and other white blood cells which are activated in these conditions. In clinical, CRP is mainly used for auxiliary diagnosis of infections, tissue injuries and inflammatory diseases.
This product should be used together with the fluoresence lateral flow immunoassay analyzer produced by Prometheus Bio Inc. It can be used in medical and healthcare institutions.

PRODUCT DETAILS

Applicable Instrument: NGFT-AC003 Prometheus SoloDx produced by Prometheus Bio Inc.
Sample: Whole blood, serum, or plasma
Reference Ranges: < 10 mg/L
Precision: Intra-batch precision CV (%) ≤ 15%; Inter-batch precision CV (%) ≤ 15%
Accuracy: Relative deviation is under ±15%
Linearity range: 0.5-150 mg/L, the correlation coefficient r ≥ 0.99
Lower Limit Detection (LLD):≤ 0.5 mg/L
No Cross-Reactivity:100 g/L albumin, 50 ng/mL Procalcitonin, 150 mg/L Serum Amyloid A, 100 ng/mL Interleukin-6, 4000 IU/ mL rheumatoid factor
Interference Tolerance:40 mg/dL bilirubin, 1000 mg/dL triglyceride, hemoglobin, cholesterol
Storage: Cool, dark and dry place at 4-30°C (Do not freeze)
Shelf-Life: 18 Months
Packaging Specification:25 kits/box, 40 kits/box, 50 kits/box, 100 kits/box
Manufacturer: Prometheus Bio Inc
RegistrationCE Mark

Additional information

Box Count

25 kits/box, 40 kits/box, 50 kits/box, 100 kits/box

Registration

TOP